EFLAB Laminar
Air Flow (LAF) cabinet instruments ensure an extremely clean environment for
product &
material/sample protection against airborne dust, particle or
microorganisms. EFLAB Vertical
LAF cabinets
work under air filtration, recirculation and indoor
exhausting principle. Room air passes through pre-filters that
block coarse particles at device top and reaches to work
cabinet area with an air velocity of 0.40 m/s (± %5) by
passing through HEPA filter. Built-in laminator in the fan
filter module supplies the airflow in a homogenous and
laminated way. Designs of working cabinet and air diffusion
assure trimmed airflow and sweeping dead zones.
Air moves in from two ways into the work chamber, one part is
exhausted to the room at the front by preventing
room air movements into the work zone. The rest of the
internal air is recirculated from the back holes of the
cabinet which increases the HEPA filter life.